- Measures ac and dc voltage, resistance, continuity w/ beeper, capacitance and frequency
- Fully autoranging with 7 measurement functions and 27 ranges
- Large oversize digital display with measurement results and the unit icon
- Data Hold freezes measurement value on
- Average sensing
- Power consumption (typical) 2 mA
- Automatic Power Off , timing Idle for 30 minutes
- APO consumption (typical) 2.2 μA
- E.M.C. Meets EN61326
- Battery installed, and User’s manual
- Audible Continuity Tester Open circuit voltage 0.4 V dc typical
- Diode Tester Test current (typical) 0.25 mA
- Diode Tester Open circuit voltage 1.6 V dc
|
|
제품사양
Warranty:
|
One-year
|
Size:
|
L 113 mm x W 53 mm x H 10.2 mm (4.4 x 2.1 x 0.4 in)
|
Weight:
|
Approx. 78 gm (3 oz)
|
Safety:
|
CAT III 300 V, CAT 600 V,
|
DC Voltage Range:
|
|
400.0 mV , Accuracy: ± (1.0 %+2 dgt)
|
DC Voltage Range:
|
|
4.000 V, 40.00 V, 400.0 V, Accuracy: ± (2.0 %+2 dgt)
|
DC Voltage Range:
|
|
600 V, Accuracy: ± (2.5 %+4 dgt)
|
DC Voltage NMRR:
|
|
> 50 dB @ 50 Hz/60 Hz
|
DC Voltage CMRR:
|
|
> 120 dB @ dc, 50 Hz/60 Hz; Rs= 1 kΩ
|
Input Impedance:
|
|
10 MΩ, 30 pF nominal; (1000 MΩ for 400.0 mV range)
|
AC Voltage Range 50 Hz - 60 Hz:
|
|
4.000 V, 40.00 V, 400.0 V, Accuracy: ± (2.0 %+5 dgt)
|
AC Voltage Range 60 Hz - 500 Hz:
|
|
4.000 V, 40.00 V, 400.0 V, Accuracy: ± (3.0 %+5 dgt)
|
AC Voltage Range 50 Hz - 500 Hz:
|
|
600 V, Accuracy: ± (3.5 %+5 dgt)
|
Input Impedance:
|
|
10 MΩ, 30 pF nominal
|
AC Voltage CMRR:
|
|
> 60 dB @ dc to 60 Hz, Rs=1 kΩ
|
Capacitance Range*:
|
|
500.0 nF, 5.000 μF, 50.00 μF,500.0 μF, 3000 μF3
|
Capacitance Accuracy*:
|
|
± (3.5 %+6 dgt4)
|
Resistance Range:
|
|
400.0 Ω, Accuracy: ± (1.5 %+6 dgt)
|
Resistance Range:
|
|
4.000 KΩ, 40.00 KΩ, 400.0 KΩ, Accuracy:± (1.0 %+4 dgt)
|
Resistance Range:
|
|
4.000 MΩ, Accuracy: ± (1.5 %+4 dgt)
|
Resistance Range:
|
|
40.00 MΩ, Accuracy: ± (2.5 %+4 dgt)
|
Open circuit voltage:
|
|
0.4 V dc typical
|
Frequency Range**:
|
|
400 Hz, 4 kHz, 40 kHz, 400 kHz, 1 MHz,Accuracy:± (0.5 %+4 dgt)
|
Sensitivity (V-peak):
|
|
5 Hz - 100 kHz, Accuracy:> 1.3 Vp
|
Sensitivity (V-peak):
|
|
100 kHz - 500 kHz, Accuracy:> 2.2 Vp
|
Sensitivity (V-peak):
|
|
500 kHz - 1 MHz, Accuracy:> 4.2 Vp
|
Audible Continuity Tester Open circuit voltage:
|
|
0.4 V dc typical
|
Audible threshold :
|
|
65 Ω ±55 Ω
|
|
|